40
41
Warning message signal 3pulses of 200ms and 150ms pause, with 2�5seconds separation sound pressure
level: min� 47 dB (A)
(measured in a hemisphere with a radius of 1 m)
Transport and storage
conditions
Temperature: -25 to +60°C
Relative Humidity: 15 to 90%, non-condensing
Pressure: 700hPa to 1060hPa
Ambient conditions in
operation
Temperature: +5 to +40°C
Relative Humidity: 15 to 90%, non-condensing
Pressure: 700hPa to 1060hPa
Dimensions H x W x D 172x214x105mm without exudate canister
Depth with 300ml exudate canister: max� 117mm
Depth with 800ml exudate canister: max� 148mm
Weight Negative pressure unit (without canister)
Exudate canister 300ml
Exudate canister 800ml
Power supply and power supply cable
1�2kg
0�25kg
0�29kg
0�50kg
Recurrent test A recurrent test must be performed every three years�
Protection class II
Degree of protection Applied parts type BF
Type of protection IP 22(Protected from touch by fingers, objects greater than 12millimetres and
water spray less than 15 degrees from vertical�)
Classification IIa (according to Annex IX EC Directive 93/42/EEC)
CE mark CE 0123
UMDNS code Suction Unit Wound 10-223
Battery Battery: Rechargeable Li-ion Battery 14,40 - 14,52V, 2A,
2600 - 2900mAh, 38,00 - 42,48Wh