35
10.2 Cleaning and disinfecting
IMPORTANT!
• Fluids in the internal parts of the device can damage the negative pressure unit and the power supply�
> The device and the power supply must never be autoclaved, rinsed under running water or immersed
in liquids�
If changing patients
> Clean the entire surface of the device using a moist (never wet) cloth�
> Following this, disinfect the device using one of the following surface disinfectants�
Please note the exposure time of the used disinfectant� After the exposure time, the unit should be dried with
a suitable cloth�
> Clean and disinfect the case, including the inserts�
> Replace and dispose of the shoulder strap and the transport bag�
If not changing patients
> Clean the entire surface of the device weekly using a moist (never wet) cloth�
> Following this, disinfect the device using one of the following surface disinfectants�
10.3 Recommended disinfectants
(Manufacturer: Bode Chemie, Hamburg, Germany)
Disinfectant Ingredients (per 100 g)
Dismozon plus (application
solution)
Magnesium monoperoxyphthalate hexahydrate 95�8 g
Kohrsolin FF (application
solution)
Glutaral 5 g
Benzyl-C12-18-alkyldimethylammonium chloride 3 g
Didecyldimethylammonium chloride 3 g
Mikrobac Tissues Benzyl-C12-18-alkyldimethylammonium chloride 0�4 g
Didecyldimethylammonium chloride 0�4 g
Bacillol 30 Sensitive Tissues /
Bacillol 30 Sensitive Foam
Ethanol 14�0 g
Propan-2-ol 10�0 g
Propan-1-ol 6�0 g
Amines, N-C10-C16-alkyltrimethylenedi-, reaction products with
chloroacetic acid
0�2 g
Bacillol zero (+)-tartaric acid 0�5 g
Sodium benzoate 0�5 g
For cleaning the negative pressure unit, all cleaning and disinfecting agents with the stated ingredients are also suitable�
The use of disinfectants containing aldehydes and amines on the same object can result in discolouration�
IFU_4095080_0302884_190824_EN.indd 35IFU_4095080_0302884_190824_EN.indd 35 19.08.24 12:3119.08.24 12:31