10. Cleaning and care instructions
10.1 Basic information
IMPORTANT!
Some disinfection solutions can cause discolouration of the plastic surfaces� Avoid the ingress of fluids
Please use disposable gloves at all times when carrying out all work�
All surface disinfectants listed in Chapter 10�3 “Recommended disinfectants”, are suitable
for disinfection�
Remove and dispose of all disposable articles, such as the exudate canister, wound dressings and
tubes, before a complete cleaning
The cleaning and disinfection measures described herein do not replace the respective local hygiene
regulations applicable for operation!
When changing the unit from one patient to another, all parts coming into contact with aspiration
material (exudate canisters, tubes and wound dressings) must be discarded�
We recommend, as a matter of principle, the documentation of all maintenance and replacement
procedures in writing
Do not use any
Disinfectants containing organic or inorganic acids or bases as these can cause corrosion damage
Disinfectants containing chloramides or phenol derivatives, as these can cause stress cracking in the
plastic materials used
CAUTION!
To avoid electric shocks, separate the power supply cable and the power supply from the negative
pressure unit and remove the mains plug before cleaning the device�
The handling of the negative pressure unit has a decisive influence on reliability and safety� The hygiene
measures below are necessary measures for protecting the patient and user from contamination and for
maintaining the functional reliability of the negative pressure unit
The cleaning and disinfection measures described here do not replace the respective regulations
applicable for operation!
Please comply with the instructions for use given by the manufacturers of the disinfectants, above
all with regard to the concentration details, information relating to material compatibility and
contact times�
Cleaning and disinfection of the negative pressure unit must be carried out according to the respective
applicable procedures for cleaning and disinfection of surfaces of other electronic, non-submersible
medical devices�
IFU_4095080_0302884_190824_EN.indd 34IFU_4095080_0302884_190824_EN.indd 34 19.08.24 12:3119.08.24 12:31