37
CAUTION!
Service is not intended during therapy�
Operational and functional faults which cannot be rectified by the measures set out in the chapter
“Rectifying malfunctions”�
Measures to be taken when sending in the negative pressure unit
If the negative pressure unit needs to be sent in after consultation with the manufacturer or an authorised service partner,
the following points must be observed:
> Send in the complete unit (seedelivery note)
> Remove all disposable materials and consumables
> Send only after thorough cleaning and disinfection
> Send in airtight packaging
> Attach a detailed description of the malfunction (see chapter Transport and storage)�
11.3 Handling of rechargeable batteries
> Always store the device with rechargeable batteries 100% charged�
> The device should never be covered, exposed to direct sunlight or charged, operated or stored in the immediate
vicinity of heaters�
> Recharge the batteries after 4weeks at the latest�
> Always recharge the batteries using the associated charging accessories�
> If rechargeable batteries are charged at a high or a low ambient temperature, their capacity cannot be fully utilised�
> If the negative pressure unit is operated at a low ambient temperature, the capacity of the rechargeable batteries
cannot be fully utilised�
CAUTION!
There is a risk of explosion if a different charger is used�
12. Accessories
Name REF
Accessories
VivanoTec Shoulder Strap 409 572
VivanoTec Bag 409 506
VivanoTec Hanger Bar 030 239
VivanoTecPro Power Supply 030 232
Power Supply Cable – Europe 030 285
Power Supply Cable – Brazil 030 286
Power Supply Cable – UK/HK 030 284