26
27
Setting the values
IMPORTANT!
Tapping on the
key closes the intermittent menu and
saves the values� If the new values are not to be saved,
wait without touching the touchscreen until the display
switches to the main menu�
7.2 Starting the therapy
CAUTION!
To achieve the correct negative pressure in the system as set, before starting the therapy ensure that all
connections are correctly made and that all therapy parameters have been set�
Select the required therapy mode
> Tap on the
key�
The negative pressure unit starts and generates the set negative pressure.
7.3 Interrupting/terminating the therapy
> Tap on the key�
The therapy is interrupted or terminated.
8.
IMPORTANT!
If no therapy has been commenced within the previous 30minutes, the inactivity warning message
appears (see chapter Device Inactive)�
Warning messages
The activities in this chapter may be carried out only by a physician or a qualified person�
If no other information is provided with the individual warning message description, the delay time for detection of
a warning message condition or for generating a warning message signal is less than 1second in each case� In the case
of discrepant information regarding the time span, the actual duration depends on the point in time of the pressure
measurement
Intermittent
Continuous
1 Warning message present
2 The warning message was suppressed
By tapping on the warning message symbol (
or ), the warning
message is displayed again.