16. Information relating to EMC (electromagnetic compatibility)
• Medical electrical devices are subject to special precautionary measures with regard to EMC and must
be installed in accordance with the EMC information described in the following�
• Portable and mobile HF communication systems can influence medical electrical devices�
• Accessories, current converters and connecting cables other than those specified can lead to an
increase in electromagnetic emission or to reduced interference immunity of the device or the system�
The VivanoTecPro negative pressure unit is intended for operation in an environment as set out below� The customer or
user of the VivanoTecPro negative pressure unit must ensure that it is operated in a corresponding environment�
16.1 Guidelines and manufacturer’s declaration – electromagnetic emissions
Emissions Tests Compliance Electromagnetic Environment - Guidelines
HF emissions in accordance with
CISPR11
Group 1 The VivanoTecPro negative pressure unit uses HF
energy exclusively for its internal functions�
Therefore its HF emissions are very low and it is
unlikely that there is any interference with nearby
electronic devices�
HF emissions in accordance with
CISPR 11
Class B The VivanoTecPro negative pressure unit is suitable for
use in all facilities including in living areas and those
directly connected to a public supply network which
also supplies buildings used for residential purposes�
Emission of harmonics in
accordance with IEC61000-3-2
Class A
Emission of voltage fluctuations/
Flicker in accordance with
IEC61000-3-3
Compliant