C-3Commissioning Ý Installing modules
Spider8
1 Commissioning
This chapter shows you the steps you need to take in order to
make your measurement chain ready for use (Spider8, PC and
transducer).
Unpack the Spider8
Check the Spider8 for damage
Is the delivery complete?
If you have the basic version of the device:
On safety grounds, all free channels (4 to 7) must be covered
with blank plates.
Adequate EMC protection cannot be guaranteed unless this is
done.
Update firmware:
Existing Spider8 can be brought up to date with the Spider8 System
CD.
Connect the existing Spider8 to the COM1 or COM2 port of the
PC with the serial cable.
Program start with Windows PC:
- Start setup
- Choose language
- Choose Spider8-firmware update
SETUP starts
If you want to install modules in channels 4 to 7:
The modules (DC/SR01 or TF/SR55; TF/SR30 - TF =
carrier-frequency) are enclosed with your device or have been
delivered separately.
Install the modules as described below.