4.2 Caster
Chapter 4: Removal, Replacement, and Adjustment Procedures
Prima™ Procedural Light Service Manual (man247ra) Page 4 - 5
DRAFT 6/4/99
4
4.2 Caster
Tools required: None
Removal
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury and equipment damage.
1. Unplug the unit from its power source.
WARNING:
When positioning the Prima™ Procedural Light to gain access to the
casters, ensure that the unit is placed in a stable position. Failure to do
so could result in personal injury or equipment damage.
2. Lay the Prima™ Procedural Light on its side, in order to gain access to the
bottom of the base (A) (see figure 4-2 on page 4-5).
Figure 4-2. Caster