2.2 Rotating Arm Malfunction
Chapter 2: Troubleshooting Procedures
Prima™ Procedural Light Service Manual (man247ra) Page 2 - 7
DRAFT 6/4/99
2
2.2 Rotating Arm Malfunction
1. Move the rotating arm to the desired position. The rotating arm retains its
position once moved.
Yes No
↓→Adjust the spring tension in the rotating arm. See “Rotating
Arm” on page 4-15. If this solves the problem, go to “Final
Actions” on page 2-4. Otherwise, call Hill-Rom Technical
Support at (800) 445-3720.
2. Go to “Final Actions” on page 2-4.