Cleaning and Care
Chapter 6: General Procedures
Prima™ Procedural Light Service Manual (man247ra) Page 6 - 3
6
Cleaning and Care
SHOCK HAZARD:
Follow the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture that would allow for liquid
pooling. Personal injury or equipment damage could occur.
WARNING:
The Prima™ Procedural Light operates at a high temperature. Allow the
unit to cool for at least 30 minutes before performing any preventive
maintenance or repair procedures. Failure to do so could result in
personal injury.
CAUTION:
Do not use harsh cleansers/detergents, such as scouring pads and
heavy duty grease removers, or solvents, such as toluene, xylene, and
acetone. Equipment damage could occur.
If there is no visible soilage with possible body fluids, we recommend that you
clean the unit with a mild detergent and warm water. If disinfection is desired,
you may use a combination cleanser/disinfectant as explained in
“Disinfecting” on page 6-4.
Steam Cleaning
Do not use any steam cleaning device on the Prima™ Procedural Light with
the exception of the sterile handle. Excessive moisture can damage
mechanisms in this unit.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom Prima Procedural Light and is the answer not in the manual?

Hill-Rom Prima Procedural Light Specifications

General IconGeneral
BrandHill-Rom
ModelPrima Procedural Light
CategoryMedical Equipment
LanguageEnglish

Related product manuals