194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 99
The Hill-Rom® 900 Accella™ bed must not be used close to or on top of other items
of equipment. If this is necessary, the Hill-Rom® 900 Accella™ bed must be tested to
confirm that it functions properly in the required configuration.
Make sure that the Hill-Rom® 900 Accella™ bed functions correctly when used in
the vicinity of other electric appliances. Mobile and portable radio frequency (RF)
communication equipment may damage the electric medical equipment.
Electric medical equipment demands special precautions regarding
electromagnetic compatibility (EMC) and must be installed and used in accordance
with the EMC-related information contained in this manual.
The use of accessories, transducers and cables other than those specified, apart
from the transducers and cables sold by the manufacturer of these devices, such as
replacements of internal components, may result in an increase and/or reduction
of the immunity of the Hill-Rom® 900 Accella™ bed.
Manufacturer's guide and declaration – electromagnetic emissions
The Hill-Rom® 900 Accella™ bed and the Wi-Fi connectivity module are designed for use in
the el
ectromagnetic environment specified below. Users must ensure that they are used in
this environment.
Emission test Compliance Electromagnetic environment - Guide
RF emissions
CISPR 11
Group 1 The Hill-Rom® 900 Accella™ bed only uses radio
el
ectric power for its internal functions. Consequently,
it only produces very weak RF emissions that are
unlikely to cause interference with nearby electronic
equipment.
CISPR 11 RF
emissions
Class A The Hill-Rom® 900 Accella™ bed can be used in all
plac
es other than domestic premises and premises
that are directly connected to the low voltage public
mains power network used to supply domestic
buildings.
Harmonic emissions
IEC 61000-3-2
Class A
Flicker
IEC 61000-3-3
Applicable
CISPR 14-1 RF
emissions
Compliant The Hill-Rom® 900 Accella™ bed is not designed to
be c
onnected to other equipment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish

Related product manuals