Page 38 194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
Graphical Caregiver Interface (GCI)™ controls
The GCI is on the foot siderail.
To activate the GCI:
Touch the screen.
To release, press the round symbol until the green
light above it comes on, then PRESS FIRMLY and slide
your finger to the right so that the remaining LEDs
come on.
When the GCI is unlocked and the screen is not touched for 1
minute, it automatically returns to the home screen. If the
screen remains untouched for a further 1 minute, the locked
screen appears. If it is not touched for another 8 minutes, the
screen switches itself off.
The screen is deactivated if the bed switches to
battery mode.
Home screen
Description
Angle of the sleep
surface
Angle of the
adjustable head
section
Bed exit alertsBackrest (Head)
angle alert
Suspend Mode
Duration alert
Access to Scale*
Access to the Bed exit
alerts
Access to the
Parameters
Home
Access to the mattress
controls*
Active therapeutic
mode
Active*
MCM*
Alerts volume
Bed exit alerts*
Position
mode: The “Patient Position” mode alert is activated when the patient starts to
move.
Exiting
mode: The “Exiting” mode alert is activated when the patient moves away from
the center of the bed to try to get out.
Out of Bed
mode: This mode must be used when the caregivers want to allow the patient
to move freely in the bed. The “Out of bed” mode alert is activated when the patient
leaves the bed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish

Related product manuals