194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 51
Mobilizing the patient
CPR
Pressing the CPR button on the caregiver keypads or
pressing the handle on the adjustable head section
activates the CPR function.
The mattress deflates in approximately 30 seconds.
Press
Rearm Matress
to restore normal operation of
the mattress.
The mattress CPR function only works if the
mattress is active and its battery is fully charged (it takes 24 hours to charge the
battery).
Transport mode
The mattress switches to transport mode (battery mode) when the bed is moved.
If the bed switches to battery mode:
• The therapeutic modes are no longer operational.
• If the mattress is in P-Max mode, the latter remains permanently active.
• A backup battery maintains the CPR function for 2 hours. When this battery is
low, the bed must be connected to the mains power supply for at least 24 hours
to fully recharge the battery. Otherwise, the CPR function cannot be
guaranteed.
• A beep sounds regularly to indicate that modes are inactive.
• If a patient whose weight is close to the SWL exits the bed, the internal pressure
of the mattress may suddenly drop, causing a low safety pressure to error to
occur. In this case, reconnect to the main power supply to reset the system.
Deflating the mattress
The mattress can be temporarily deflated without stopping it.
1. Press Mattress on the CGI.
2. Press Deflate.
3.
The corresponding control turns green.
.
4. To switch off the deflation mode, press
Deflate or the Boost™ button.
5. The mattress switches to initialization mode.
After 20 minutes, the system automatically returns to
the initial therapeutic mode in order to reduce the
risks incurred in the non-therapeutic mode.