Page 80 194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
Syringe-driver holder (AC963A)**
Safe working load: 15 kg
(1)
Do not position the accessory facing inwards, particularly under the head section
when it is raised, so as to prevent any risk of the accessory obstructing the head
section or siderail when being handled.
This accessory is designed to accept a syringe-driver and is fitted at the head end
of the bed in the sockets provided.
To adjust position of the syringe driver holder:
hold the tablet and loosen the knob,
position the tablet as required and then tighten the knob.
3
1
2
Traction frame
Using traction frames on beds fitted with a weigh system compromises the
accuracy of the results of the weighing operation and of the bed exit alert
detection.
Traction equipment can be installed at four points: two at the head end and two at
th
e foot end.
The caregivers must assess the patients to prevent them from becoming trapped or
being asphyxiated when using the traction equipment.
Follow the establishment’s protocol applying to the disabling of the bed controls
when installing the traction equipment. This will avoid any injuries.
1. The safe working load specifications allow for a substantial safety margin.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish

Related product manuals