Page 50 194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
Therapeutic mode
Continuous low pressure (CLP) mode
Press CLP to select this mode.
The corresponding control turns green.
The patient is supported at an optimal low pressure controlled by
the I-mmersion™ sensor.
Alternating low pressure (ALP) mode
Press ALP to select this mode..
The corresponding control turns green.
The patient is supported at an optimal low pressure controlled by
the I-mmersion™ sensor. The cushions deflate alternately in a
complete cycle lasting about 10 minutes.
The therapeutic modes are no longer operational when the bed switches to battery
mode.
Maximum inflation mode (P-Max)
Press P-Max
on the GCI, or briefly press the Boost™ button on the
caregiver keypad to select this mode.
The corresponding control turns green.
After 20 minutes, the system automatically returns to the initial
therapeutic mode in order to reduce the risks incurred in the non-
therapeutic mode.
After activating the P-Max mode, it is possible to return to the previously
selected therapeutic mode by pressing P-Max, ALP ou CLP.
If the bed switches to battery mode when the mattress is in P-Max mode, the latter
remains active permanently. In this case, it is necessary to connect the bed to the
mains power supply to return the mattress to a therapeutic mode. Remaining in
P-Max mode for a lengthy period can result in discomfort for the patient.
MCM™*
The microclimate management system is activated automatically when the
Accella™ Therapy system is switched on. The corresponding control turns
green.
It can be deactivated by pressing MCM.