Page 72 194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
CPR
Never allow a non-qualified person to operate this function and check that no
obstacles (e.g., limbs, accessories, objects, power cords) or persons are under the
head section.
This function is used in emergencies (e.g., reanimation, heart massage, etc.) or in the
event of a power failure.
On the Hill-Rom® 900 Accella™ bed, there are two ways to operate the CPR function:
CPR handle
Pressing the handle mechanically returns the adjustable head section to the flat
position, automatically aligns the HiLow with the higher side and deflates the
combined mattress*.
As soon as the head section is flat, the handle can be released and the automatic leveling
continues (unless another movement is activated).
CPR button
Press and hold the CPR button on the caregiver half-siderail keypads.
All the sections of the sleep surface return to the flat position, the bed
frame is aligned with the lower side and the combined mattress*
deflates.
NOTE:
To reset the mattress, see “CPR” page 51.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish

Related product manuals