194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 41
Mobilizing the patient
Backrest (Head) angle alert
The
Backrest (Head) angle
alert allows the caregiver to program a sound
alert when the angle of the head section is less than 30° or 45°. A message
appears on the GCI when the head section descends below the selected angle.
This information can be sent to the duty nurse by a compatible Wi-Fi system.
The adjustable head section angle alert is longer operational when the bed
switches to battery mode.
To activate this function
1. Put the adjustable head section in the
required position, at an angle greater
than 30° or 45°.
2.
Press Alerts on the GCI.
3. Set the Backrest (Head) angle detection to the required angle:
less
than 30° or 45°.
To deactivate the
Backrest (Head) angle
detection
1. Press Alerts on the GCI.
2. Press
detection active.
3. Yes. The Backrest (Head) angle detection is deactivated.
When an alert sounds
When the adjustable head section detection is on and it
detects an alert condition, an alert signal sounds, an alert
message appears on the GCI and the night light comes on.
Press the appropriate button to switch off the night
light (see “Night light” page 71).
Raise the section to an angle greater than 30° or 45°.
or,
Press
Suspend Alert for 30 sec
on the touch screen, then select the alert deactivation
mode (see “Selecting alert suspension mode” page 42).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish

Related product manuals