194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 43
Mobilizing the patient
“Awaiting patient” mode
In this mode, the Bed exit alerts (Bed exit alerts)
are deactivated until the patient returns.
IThe monitoring system can be completely
deactivated by pressing Disarm.
When the patient returns to the bed, the system
reactivates the alerts.
If the system is unable to reactivate itself after a
given length of time, the Bed exit alert sounds.
Scale (weigh system)*
In the Scales screen of the GCI, you can reset the scales, weigh in a range from 10 to 250
kg with a resolution of 500 g in all positions authorized by the system (+/-2° from the
horizontal), increase the resolution to 100 g temporarily and display a difference in weight
in comparison with an initial weight and the patient’s BMI.
Description of the scale screen
Add/remove parts
Access to the weight
difference data
Information on the weighed /
unweighed zones
Weight
measured
Indication of the
stability of the
weight
Date of the last
reset/tare operation
Temporary display at
100 g
Reset/Tare
New patient
Before installing a new patient, it is necessary to erase the preceding patient’s data.
1. Press Parameters on the CGI.
2. Press New Patient.
This function deletes the history, deactivates the various alerts
and performs a reset/tare operation.