Page 102 194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
IMMUNITY to Proximity Fields from Radio Frequency Wireless
Communications Equipment
In addition to the Radiated RF IEC 61000-4-3 as shown in the table above, the Wi-
Fi connectivity module has been tested as specified in the table below.
Test
Frequency (MHz)
Band (MHz)
Service
Modulation
Maximum
Power (W)
Distance (m)
Immunity Test
Level (V/m)
385 380 - 390 TETRA 400 Pulse modulation
18 Hz
1,8 0,3 27
450 430 - 470 GMRS 460,
FRS460
FM +
5 kHz deviation
1 kHz sine
20,328
710 704 - 787 LTE Band 13, 17 Pulse modulation
217 Hz
0,2 0,3 9
745
780
810 800 - 960 GSM 800/900
TETRA 800, iDEN
820, CDMA 850,
Band LTE 5
Pulse modulation
18 Hz
20,328
870
930
1720 1700 - 1990 GSM 1800 ;
CDMA 1900 ;
GSM 1900 ; DECT ;
Band LTE 1,3, 4,
25 ; UMTS
Pulse modulation
217 Hz
20,328
1845
1970
5240 5100 - 5800 WLAN
802.11
a/n
Pulse modulation
217 Hz
0,2 0,3 9

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish

Related product manuals