194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 33
Mobilizing the patient
Before using this function, check that:
• the bed frame extension is securely locked in one of the notches and that
nothing (e.g., objects, accessories, power cables, tubes) or persons
(especially children) are under the sleep surface.
• the patient’s limbs are within the sleep surface,
• there is enough space between the head of the bed and the partition,
especially for Trendelenburg,
• no accessories (IV pole in particular) may come into contact with the
fittings
,
• check that the drainage devices do not come into contact with the floor.
Electrical Trendelenburg/Reverse Trendelenburg
The electrical Trendelenburg / Reverse Trendelenburg is operated using the
caregiver controls on the half-siderail.
Before using this function, check that the HiLow is enabled.
To tilt the sleep surface:
• press the required function (A) or (B) at the same time,
• release the button when the req
uired angle is attained.
This function can be used without a mains power supply thanks to the battery.
Boost™ patient position mechanism
Before using this function, check that the HiLow is enabled.
This function puts the sleep surface sections in the flat position and inclines it into the
Trendelenburg position to 7°. It also activates the P-Max mode of the combined mattress*.
To activate this function, press and hold the Boost™ button and release when at the
required angle.
This function can be used without a mains power supply thanks to the battery.
Briefly pressing the Boost™ button activates the P-Max function of the mattress
(see “Maximum inflation mode (
P-Max)” page 50).