194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 45
Mobilizing the patient
4. Check that neither the mattress nor any accessories are touching the fixed parts
of the bed (and in particular the head section) and that no traction is applied to
the parts installed above and below the sleep surface (e.g., power cable and air
mattress pipes). If necessary, remove the headboard.
Correct position
Incorrect position
Headboard
Zero/Tare
1. Press Scale on the GCI.
2. Press Zero/Tare.
3. Follow the instructions.
– If a message appears on the GCI
wh
en resetting, adjust the bed
accordingly.
4. Reset/tare operation complete:
– Deletes the patient weight
diff
erence data and
resets the weigh
system.
– The screen shows 0.0 kg and the
>
0< light comes on to show the accuracy of the reset/tare operation.