194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use Page 59
Mobilizing the patient
Installation
1. Initialization: When the device is connected to the interface board, the ON/OFF
indicator light is red and the other indicator lights remain off for about 8 seconds
2. The ON/OFF indicator light is red and the other indicator lights flash red, then
green, then go off, and so on, during the initialization cycle.
3. The bed recognizes the device. The ON/OFF indicator light is red and the other
lights are off.
4. Connection to the wireless network: Press the ON/OFF button. The ON/OFF
light turns green and the other lights are red.
5. Wi-Fi configuration of the MCW. It is necessary to update the Wi-Fi settings of the
MCW in order to activate the device. Use the Hill-Rom LI900B4 Tool software to
do this (see manual 209584).
Only qualified and authorized technician should perform this operation
Use
WCM Startup
1. Press ON/OFF button.
2. The ON/OFF light turns green
3. Connecting to Wi-Fi facility network: The Wi-Fi light turns green.
4. Connecting to the server facility: The Server Connection light turns green.
5. Bed Location: The Bed Location light turns green.
NOTE:
The bed is located automatically if it is connected to SmartSync™, and manually if it is
connected to the
Hillrom™ Digital Health Gateway.
The bed can then transmit the information to the facility communication system
and the bed is located and associated with the room.
When moving the bed outside the room (battery mode), the WCM will turn off.
After reconnecting the bed to the mains, make sure that all lights are green. If
location light remains red, check that the room is equipped with a tracking
system and is operational.
The Wi-Fi system does not replace the visual and audible alerts generated by the
bed. Caregivers should use these local alerts and bed information to make a
decision.
Stopping the WCM
1. Press ON/OFF button.
2. The ON/OFF light is red and the other lights are off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish

Related product manuals