Accella™ Therapy Mattress Instructions for Use
RF communication equipment (including peripherals such as antenna cables and external
antennas) should be used no closer to any part of the device, including cables specified by
Hill-Rom, than the recommended separation distances shown in the following table.
Otherwise, degradation of the performance of this equipment could result.
*:Effective Radiated Power.
Wireless communication system Separation
distance
(m)
TETRA 400 (limited to 10 W ERP*) 0.3
Public safety (460-470 MHz), GMRS 460 (limited to 5 W ERP*) 0.2
GMRS 460 (limited to 2 W ERP*) 0.1
GSM 850, GSM 900, RFID 868 MHz, TETRA 800 (limited to 2 W ERP*) 0.3
GSM 1900 (limited to 1 W ERP*) 0.2
WLAN 802.11a 5 GHz (limited to 1 W ERP*) 0.7
iDEN 820, CDMA 850, GSM 1800, CDMA 1900 (limited to 0.6 W ERP*) 0.2
FRS 460 (limited to 0.6 W EPR*), PMR 446 0.1
UMTS, DECT (limited to 0.25 W EPR*) 0.1
Bluetooth, WLAN 802.11 b/g 2450 , RFID 2450 (limited to 0.1 W EPR*) 0.1