199253
(6) -
EN
-
Accella™ Therapy Mattress Instructions for Use Page 57
Compliance
Table 6 – Recommended separation distances between portable and mobile RF
communication equipment and Accella™ Therapy device –for the Accella™ Therapy device
Recommended separation distances between portable and mobile RF communications
equipment and the
Accella™ Therapy
The Accella™ Therapy device is designed for use in an electromagnetic environment in which
interference due to radiated RF is monitored. The customer or user of the Accella™ Therapy
device can contribute to the prevention of electromagnetic interference by keeping the
Accella™ Therapy device at the recommended distances from portable and mobile RF
equipment (transmitters) as shown below, according to the maximum power output of the
communication equipment.
Maximum
transmitting power of
the transmitter
W
Separation distance versus the frequency of the transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5
GHz
0.01
0.12 0.12
0.23
0.1
0.37 0.37
0.74
1
1.17 1.17
2.33
10
3.69 3.69
7.39
100
11.67 11.67
23.33
For transmitters with a maximum transmitting power not given above, the recommended
separation distance in meters (m) can be estimated by using the equation applied to the
transmitter frequency, where P is the maximum transmitting power of the transmitter in Watts
(W), according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the highest frequency range
applies.
NOTE 2: These directives cannot be applied to every situation. The propagation of
electromagnetic waves is affected by absorption and reflection due to structures, objects and
persons.