8.6 Message Reference
124
Set and Query the Auto-Ranging Setting
Syntax
Command
:AUTorange <1, 0, ON or OFF>
Query
:AUTorange?
Response <ON or OFF>
Example
Command
:AUT ON
Note
Attempting to enable auto-ranging when the Comparator or Memory function is
enabled results in a execution error.
The auto-ranging setting applies to both resistance measurement and voltage
measurement.
Cancel Zero-Adjustment
Syntax
Command
:ADJust:CLEAr
Description
Command Clears zero adjustment.
Execute Zero Adjustment and Query the Result
Syntax
Query
:ADJust?
Response <0/ 1 (NR1)>
0 ....... Zero adjustment succeeded
1 ....... Zero adjustment failed
The acceptable range of zero adjustment for both resistance
and voltage is
± 1000 dgt.
Description
Query Queries whether zero adjustment has succeeded or failed.
Example
Query
Response
:ADJ?
0
Zero adjustment executed successfully.
Select and Query the Sampling Rate setting
Syntax
Command
:SAMPle:RATE <EXFast/ FAST/ MEDium/ SLOW>
Query
:SAMPle:RATE?
Response <EXFAST/ FAST/ MEDIUM/ SLOW>
Example
Command
:SAMP:RATE MED
Query
Response
:SAMP:RATE?
MEDIUM