3.2 Basic Measurement Example
24
9
Set the sampling rate. (Here, SLOW is selected.)
See "3.5 Setting Sampling Rate" (⇒ p.30).
The sampling rate changes each time you
press this key.
EX.FAST
FAST
SLOW
MED
SLOW lit
Zero-Adjustment
10
Short the test leads together.
Proper Zero-Adjustment is not possible with incorrect wiring.
See "3.6 Zero-Adjust Function" (⇒ p.31).
Example: Model 9770 Pin Type Lead
11
(The SHIFT indicator lights up.)
Execute Zero-Adjust.
After zero-adjustment, the display returns to the measurement mode.
“Err.02” appears if Zero-Adjustment fails. Verify that the test lead tips are properly shorted,
and try zero-adjustment again.
Bring the pins into contact at 3 points.
• Internal conductor and internal conductor
• Internal conductor and external conductor
• External conductor and external conductor
Model 9770
Internal
conductor
External
conductor