9.3 General Specifications
177
9
Chapter 9 Specifications
Resistance Measurement
*1: Measurement current error within
±10%
*2: Add
± 3 dgt for EX.FAST, or ± 2 dgt for FAST and MEDIUM sampling rates.
: Add
± 30 dgt for EX.FAST, or ± 10 dgt for FAST, or ± 5 dgt for MEDIUM sampling rates 3mΩ range.
Voltage Measurement
*3: Model BT3563 only
*4: Add
± 3 dgt for EX.FAST, or ± 2 dgt for FAST and MEDIUM sampling rates.
Range 3 mΩ 30 mΩ 300 mΩ 3 Ω 30 Ω 300 Ω 3000 Ω
Maximum displayed
values
3.1000 mΩ 31.000m Ω 310.00 mΩ 3.1000 Ω 31.000 Ω 310.00 Ω
3100. 0 Ω
Resolution
0.1
μΩ 1 μΩ 10 μΩ 100 μΩ
1 m Ω 10 mΩ 100 mΩ
Measured current
*1
100 mA 100 mA 10 mA 1 mA
100
μA 10 μA10 μA
Measured current
frequency
1 kHz
± 0.2 Hz
Accuracy
*2
± 0.5%rdg. ± 5dgt.
± 0.5%rdg. ± 10dgt. (3 mΩ)
Temperature coeffi-
cient
(
± 0.05%rdg. ± 0.5dgt.)/°C
(
± 0.05%rdg. ± 1dgt.)/°C (3 mΩ)
Range 6 V 60 V
300 V
*3
Maximum displayed
values
± 6.00000 V ± 60.0000 V ± 300.000 V
Resolution
10
μV 100 μV
1 mV
Accuracy
*4
± 0.01%rdg. ± 3dgt.
Temperature coeffi-
cient
(
± 0.001%rdg. ± 0.3dgt.)/°C
9.3 General Specifications
Operating temperature and
humidity
0 to 40°C (32 ± 104°F), 80%RH or less (non-condensating)
Storage temperature and humidity -10 to 50°C (14 ± 122°F), 80%RH or less (non-condensating)
Temperature and humidity range
for guaranteed accuracy
Period of guaranteed accuracy
23 ± 5°C (73 ± 9°F), 80%RH or less (non-condensating)
1 year
Operating environment Indoors, Up to 2000 m (6562 ft) ASL
Rated supply voltage AC100 V to AC240 V (Figures reflect assumed voltage fluctuations
of 10%.) (Auto selecting), anticipated transient overvoltage 2500 V
Rated supply frequency 50 Hz/ 60 Hz
Power consumption 30 VA