3.2 Basic Measurement Example
23
3
Chapter 3 Measurement
Instrument Settings
5
Confirm the SHIFT lamp is not lit.
If this is lit, press the SHIFT key to turn it off.
6
Select the Resistance Measurement mode.
(Here, resistance and voltage measurement is selected.)
See "3.3 Selecting Measurement Mode" (⇒ p.26).
7
Set the measurement range. (Here, 30 mΩ range is selected.)
See "3.4 Setting Measurement Range" (⇒ p.27).
8
(SHIFT Lamp lit)
Set the voltage measurement range. (Here, the 6 V setting has been selected.)
See "Voltage measurement range" (⇒ p.28)
The measurement mode changes each time
you press this key. Select Ω to measure only
resistance, or V to measure only voltage.
ΩV
Ω
V
ΩV lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
mΩ lit
Increase the voltage measurement range.
Decrease the voltage measurement range.
V lit