3.4 Setting Measurement Range
27
3
Chapter 3 Measurement
This section describes how to set the measurement range for resistance or voltage measurement. For resis-
tance measurement, you can select from seven ranges from 3 mΩ to 3000 Ω. For voltage measurement, you
can select from two ranges from 6 V to 60 V (for the Model BT3562[-01]) or three ranges from 6 V to 300 V
(for the Model BT3563[-01]). There is also an auto-range function, which determines the optimal range auto-
matically.
3.4 Setting Measurement Range
Resistance measurement range
1
Select the resistance measurement range.
The position of the decimal point and unit of measurement on the display will
be switched according to the selected range.
2
Select the range to use.
Ω lit
mΩ lit
When the 3 Ω range is selected
When the 300 m
Ω range is
selected
Increase the resistance measurement range.
Decrease the resistance measurement range.
Pressing the up or down keys while in auto-range mode will cancel auto-ranging,
leaving the current measurement range as the manually set range.