3.4 Setting Measurement Range
29
3
Chapter 3 Measurement
*1: BT3563 only
Auto-Ranging
When manual range selection is enabled, pressing this enables auto-ranging.
The most suitable measurement range is then selected automatically.
The auto-range setting (on/off) for the ΩV function applies to both resistance and
voltage measurement.
Switching from
Auto
-ranging back to
Manual
range selection
Press the AUTO key again. The range can now be changed manually.
• Depending on the state of the test object, auto-ranging may be unstable. In
this case, select the range manually, or increase the Delay time.
• Auto-ranging is not available when Comparator or Memory functions are
enabled (ON).
• Refer to "Specifications" (⇒ p.171) for details about accuracy.
Range Displayed Values
Resistance Measurement Mode
Measured Current Open-Terminal Voltage
3 mΩ -0.1000 to 3.1000 mΩ 100 mA 25 Vpeak
30 mΩ -1.000 to 31.000 mΩ 100 mA 25 Vpeak
300 mΩ -10.00 to 310.00 mΩ 10 mA 7 Vpeak
3 Ω -0.1000 to 3.1000 Ω 1 mA 4 Vpeak
30 Ω -1.000 to 31.000 Ω 100 μA 4 Vpeak
300 Ω -10.00 to 310.00 Ω 10 μA 4 Vpeak
3000 Ω -100.0 to 3100.0 kΩ 10 μA 4 Vpeak
6 V -6.00000 V to ±6.00000 V -- --
60 V -60.0000 V to ±60.0000 V -- --
300 V
*1
-300.000 V to ±300.000 V -- --