4.1 Comparator Function
52
2
Display the absolute value judgment function configuration screen.
See "1.4 Menu Display Sequence (SHIFT ENTER)" ( p.13)
3
Set the absolute value judgment function to either "On" or "Off."
on
.........................
Absolute value judgment function on.
oFF................... Absolute value judgment function off.
4
Accept the setting and return to the measurement screen.
For example, the following judgment results would be obtained when connecting
the probes backwards to a 3.7 V battery (resulting in a displayed voltage mea-
sured value of -3.7 V) with an upper threshold of 3.9 V and a lower threshold of
3.6 V:
Absolute value judgment function off: Lo
Absolute value judgment function on: IN
Enabling and Disabling the Comparator Function
Enables the comparator
When the comparator is enabled, the following key operations are disabled
to avoid inadvertent operations.
ΩV/Ω/V key (Measurement mode setting)
SHIFT
ΩV/Ω/V key (Zero-Adjustment)
SHIFT
COMP key (Comparator setting)
AUTO key (Auto-ranging setting)
SMPL key (Sampling rate setting)
SHIFT
SMPL key (Averaging setting)
SHIFT
TRIG key (Trigger source setting)
SHIFT
ENTER key (Menu display)
SHIFT
STAT key (Delay setting)
Range keys
COMP lit
When the comparator is enabled, auto-ranging is automatically disabled.