4.11 Measurement Value Output Function
70
This function causes output of measured values via the RS-232C interface in the
same sequence as trigger input.
This function is useful when measuring using internal (free-run) triggering, and
for obtaining measured values on a PC when using a footswitch for triggering.
4.11 Measurement Value Output Function
1
(The SHIFT indicator lights up.)
The Menu display appears.
2
The Measurement Value Output function setting display appears.
See "1.4 Menu Display Sequence (SHIFT → ENTER)" (⇒ p.13).
3
Turn Measurement Value Output Function on or off.
on.......... enables the measurement value output function on.
oFF........ disables the measurement value output function off.
4
Applies setting and returns to the Measurement display.
5
The measured value is output from the RS-232C interface when you
press the
TRIG key or when a signal is applied to the EXT I/O TRIG
terminal.
Set the PC to the receiving state beforehand. When a measurement
value is received, the PC should perform appropriate processing
such as recording or displaying, then re-enable the receiving state.
(Main Display)
(Sub Display)
The current setting blinks.
• When external triggering is enabled, a measurement is performed and the
value is sent after each trigger event. When internal triggering is enabled,
the first value measured after triggering is sent.
• The measurement output function is not applicable to the GP-IB interface or
printer.