8.6 Message Reference
129
8
Chapter 8 RS-232C/GP-IB Interfaces
Set and Query the Comparator Reference Value
(Resistance Measurement)
Syntax
Command
:CALCulate:LIMit:RESistance:REFerence
<Reference value>
Query
:CALCulate:LIMit:RESistance:REFerence?
Response <Reference value>
<Reference value> = 0 to 99999 (NR1)
Example
Command
:CALC:LIM:RES:REF 5076
Sets the reference value to 50.76 mΩ (with the 300 mΩ range
selected)
(If the 3
Ω range is selected, the threshold is set to 0.5076 Ω)
Query
:CALC:LIM:RES:REF?
Response
5076
Note The value is sent as a whole integer (count). To set 120.53 mΩ with the 300 mΩ
range, send the following:
:CALC:LIM:RES:REF 12053
(Voltage Measurement)
Syntax
Command
:CALCulate:LIMit:VOLTage:REFerence
<Reference value>
Query
:CALCulate:LIMit:VOLTage:REFerence?
Response <Reference value>
<Reference value> = 0 to 999999 (NR1)
Example
Command
:CALC:LIM:VOLT:REF 370000
Sets the reference value to 3.70000 V. (with the 6 V range selected)
Query
:CALC:LIM:VOLT:REF?
Response
370000
Note
The value is sent as a whole integer (count).
To set 47.0000 V with the 60 V range, send the following:
:CALC:LIM:VOLT:REF 470000