8.6 Message Reference
133
8
Chapter 8 RS-232C/GP-IB Interfaces
Query the Data Count
(Resistance Measurement)
Syntax
Query
:CALCulate:STATistics:RESistance:NUMBer?
Response <Total data count (NR1)>,<Valid data count (NR1)>
<Total data count (NR1)> = 0 to 30000 (NR1)
<Valid data count (NR1)> = 0 to 30000 (NR1)
Example
Query
:CALC:STAT:RES:NUMB?
Response
22,20
Note
Measurement faults and out-of-range "OF" measurements are ignored for statisti-
cal calculations.
(Voltage Measurement)
Syntax
Query
:CALCulate:STATistics:VOLTage:NUMBer?
Response <Total data count (NR1)>,<Valid data count (NR1)>
Example
Query
:CALC:STAT:VOLT:NUMB?
Response
22,20
Note
Measurement faults and out-of-range "OF" measurements are ignored for statisti-
cal calculations.
Query the Mean value
(Resistance Measurement)
Syntax
Query
:CALCulate:STATistics:RESistance:MEAN?
Response <Mean (NR3)>
Example
Query
:CALC:STAT:RES:MEAN?
Response
295.76E-3
(Voltage Measurement)
Syntax
Query
:CALCulate:STATistics:VOLTage:MEAN?
Response <Mean (NR3)>
Example
Query
:CALC:STAT:VOLT:MEAN?
Response
1.3923E+0