8.6 Message Reference
153
8
Chapter 8 RS-232C/GP-IB Interfaces
:COMParator
:COMParator?
0 to 30 Range of panel numbers:
Turns Off when the panel num-
ber is 0, and turns On when the
panel number is 1 to 30
Response:
Returns 0 when the compara-
tor is disabled (OFF), and 1
when enabled (ON)
Range of Comparator Num-
bers: 0 to 30
Response:
Returns the response number
:CSET:MODe
:CSET:MODe?
R/ RV
:CSET:NUMBer
:CSET:NUMBer?
1 to 126 (function not available) Specifies the comparator table
number to set
:CSET:RPARameter
:CSET:RPARameter?
<Upper threshold/ Lower
threshold>
Setting range:
0 to 3.1000E+3
*Be sure to set the measure-
ment range first. Otherwise,
this setting will not be properly
configured.
Setting range: 0 to 3.1000E+3
:CSET:RRANge
:CSET:RRANge?
0 to 3E+0
3E-3 to 3E+3
Resistance range: 0 to 3.1E+3
3 mΩ ranges are supported.
Resistance range: 0 to 3.1E+3
:CSET:VPARameter
:CSET:VPARameter?
<Upper threshold/ Lower
threshold>
Setting range: 0 to 300 V
* Negative setting values are
invalid.
*Be sure to set the measure-
ment range first. Otherwise,
this setting will not be properly
configured.
Setting range:
-5.0000 to 5.0000
(5 V range)
-50.000 to 50.000
(50 V range)
:CSET:VRANge
:CSET:VRANge?
-300 to 300
6E+0/60E+0/300E+0
Voltage range: -300 to 300
6 V, 60 V, and 300 V ranges
are supported.
*The 300 V range is only avail-
able on the Model BT3563(-01).
Voltage range: -50 to 50
Response: 5E+0/ 50E+0
:CTMode
:CTMode?
AUTo/ MANual
:MEASure:BATTery?
<Measured resistance,
Measured voltage, Judg-
ment result>
FAIL/ PASS/ OFF/ NG
Resistance measurement val-
ues consist of five digits with
FAST sampling
Voltage measured values: 1
digit for sign + 6 digits for value
* Numerical values do not in-
clude a decimal point.
Resistance measurement val-
ues consist of four digits with
FAST sampling
Voltage measured values: 1
digit for sign + 5 digits for value
* Numerical values do not in-
clude a decimal point.
:MEASure:RESistance?
<Measured resistance,
Judgment result> FAIL/
PASS/ OFF/ NG (ΩV)
HI/ IN/ LO/ OFF/ NG (Ω
)
Resistance measurement val-
ues consist of five digits with
FAST sampling
* Numerical values do not in-
clude a decimal point.
Resistance measurement val-
ues consist of four digits with
FAST sampling
* Numerical values do not in-
clude a decimal point.
:MEASure:VOLTage?
<Measured voltage, Judg-
ment result> FAIL/ PASS/
OFF/ NG
Response:
Mark: one character + six nu-
merals
* Numerical values do not in-
clude a decimal point.
Response:
* Numerical values do not in-
clude a decimal point.
:FREQuency
:FREQuency?
AUTO/50/60 Setting range: AUTO/50/60
Power supply frequency set-
ting: Support for AUTO detec-
tion
Setting range: 50/60
Message ([ ] = optional) Data Contents
( ) = response data
Differences
Model BT3562 (BT3562-01) Model 3560

Table of Contents

Related product manuals