9.1 Basic Specifications
174
Comparator setting Comparator execution mode: HIGH, LOW/REF, %
Upper and lower threshold: 0 to 99999 (Resistance)/ 0 to 999999 (Voltage)
Reference value and tolerance:0 to 99999 (Resistance)/ 0 to 999999
(Voltage)
%: 0.000% to 99.999% (percentage range setting applies to both positive
and negative values)
Comparator judgment beeper: OFF/ HIGH, LOW/ IN/ ALL
• Operating mode: AUTO/ MANUAL
* Measurement value or statistical 3σ (population standard deviation ×3)
can be set as upper threshold or reference values.
Decision Judgment result: Hi/ IN/ Lo (resistance and voltage judged independently)
Calculates the logical AND of resistance and voltage judgment results and
outputs a PASS/FAIL judgment (EXT I/O output).
Measurement fault value judgments:
OF Hi judgment
-OF Lo judgment
Measurement fault Not judged (no judgment result)
*Configuring the Absolute Value Judgment Function (Voltage)
Comparator
Statistical Calculation
Statistical calculation ON/ OFF/ clear
Auto-clear after printing statistical data
Calculations Total data counts, Valid data counts, Maximum, Minimum, Mean, Standard
deviation, Population standard deviation and Process capability indices
(Cp and CpK)
Calculations trigger Statistical calculation of measured values initiated by EXT I/O signals, key
or remote command
Measurement Memory and Batch Download Functions
Measurement memory ON/ OFF/ clear
Memory trigger Up to 400 measurement values can be stored in internal memory by EXT I/
O signals, key or remote command.
Stored measurement values can be batch downloaded by remote com-
mand.
*Data stored in memory cannot be displayed on the instrument.
Key-Lock
Key-lock ON/ OFF
Key operations are disabled when ON.
Power supply frequency
Power supply frequency
setting
AUTO (automatic selection of 50 Hz/60 Hz) /50 Hz/60 Hz