EasyManua.ls Logo

Hioki BT3562-01 - Chapter 4 Applied Measurement

Hioki BT3562-01
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.6 Zero-Adjust Function
32
2
Short the test leads together.
Proper zero adjustment is not possible with incorrect wiring.
Example: Model L2107 CLIP TYPE LEAD
Red
SOURCE
SOURCE
SOURCE
SENSE
SENSE
SENSE
Correct
Incorrect
Red
Black
Black
SOURCE
SENSE
Bring the "V" marks together at the
same position.
Model 9770 (Option)
Model 9453 (Option)
Perform zero adjustment with
the alligator clips and lead
rods placed as above.
External conductor
Bring the pins into
contact at 3 points.
Model L2100 (Option)
Model 9454
Zero Adjustment Board
Line
Each sensor pin has a line affixed to its base. When
using the zero-adjust feature, align these lines in the
same direction.
As shown in the illustration on the right, select a hole
suited to the distance between terminals on the bat-
tery subject to measurement. Press it in a way sym-
metrical to the central screw on the zero-adjust board.
Insert the sensor pin (line side) into the hole.
Model 9771 (Option)
Let the two points of the pin tip touch the spring
part perpendicularly (be careful not to short the
springs).
When the resistance measurement value is dis-
played as "-----", change the facing direction.
Internal
conductor

Table of Contents