4.3 Measurement Current Pulse Output Function
58
Measurement current application timing when measurement current pulse mode is on
Measurement processing triggered by external TRIG signal or "READ?" command
Applied only during measurement (including trigger delay time).
Measurement current is applied continuously.
Measurement and current application when using external trigger measurement
or when continuous measurement is off (:INIT:CONT OFF)
Measurement current application when using normal internal trigger measurement
(The measurement current is applied continuously when measurement current pulse mode is off.)
Measurement
Measurement
Measurement
Measurement
current
current
When the measurement current pulse output function is on, the below is added
to the sampling time.
When using the ΩV mode: +1 ms
When using the Ω mode: +4 ms