General Specications
Dimensions Approx. 93W × 197H × 53D mm
(3.66″W × 7.76″H × 2.09″D) (excluding the projection)
Mass Approx. 650 g (22.9 oz.) (including batteries)
Applicable
standards
• Safety: EN 61010
• EMC: EN 61326
• Dustproof and waterproof: IP40 (EN 60529)
IMPORTANT
To avoid any failure, do not allow the
instrument to get wet. If the instrument gets
wet, have your authorized Hioki distributor or
reseller inspect or repair it, if necessary.
Accessories • L9207-10 Test Lead
• Instruction Manual
• LR6 alkaline battery × 4 (unloaded)
Replacement parts • µA/mA terminal fuse (DT4281, DT4282)
Manufacturer HOLLYLAND
Rating 630 mA/1000 V
Breaking
characteristics
Fast-blow type
Breaking capacity 50 kA AC/30 kA DC
Size
φ
10.3 mm × 38 mm
Resistance Approx. 1.2
Ω
Printed color Blue
• A terminal fuse (DT4282 only)
Manufacturer HOLLYLAND
Rating 11 A/1000 V
Breaking
characteristics
Fast-blow type
Breaking capacity 50 kA AC/30 kA DC
Size
φ
10.3 mm × 38 mm
Resistance 0.1
Ω
or less
Printed color Red
Options See “Options (sold separately)” (p. 2).