24
Alarm Display and Battery Indicator
When the Thermocouple (K) is broken (Temperature
measurement)
TEMP C
TEMP FVERSION
RESET
0-20mA電源遮断断線(K熱電対)####
オートパワーセーブOFFブザーOFFメモリデータクリアメモリデータなし
Corrective action:
Check that the thermocouple has been connected
correctly to the measurement terminal. If the display
does not change, replace with a new Thermocouple
(K). (p. 45)
Battery indicator
Fully charged. (Charge: 60% or more)
As the battery charge diminishes,
black charge bars disappear, one
by one, from the left of the battery
indicator.
(Charge: 20% or more)
The battery voltage is low. Replace the
batteries as soon as possible.
(Charge: 5% or more)
(Blinks) The battery is exhausted.
Replace the batteries.
(Charge: less than 5%)
The charge is only a reference for the continuous operation time. (p. 96)
Power shutdown
TEMP C
TEMP FVERSION
RESET
0-20mA電源遮断断線(K熱電対)####
オートパワーセーブOFFブザーOFFメモリデータクリアメモリデータなし
When the charge is 0% (less than 3.8 V ± 0.2 V),
[bAtt Lo] is displayed for 1 second and the power is
shut down.