58
Retaining the Measured Value
• If the input signal is too small for the relevant range (dead zone
threshold p. 58), the measured value cannot be automatically
retained.
• If the difference between the current and previous measured
values remains less than the counts in the stable eld of the
table below, the instrument determines the current value has
stabilized.
Dead zone threshold
Function Stable (count)
Dead zone threshold
(count)
AC voltage
1200, 200
(1000 V range)
1200, 200
(1000 V range) *
DC voltage
1200, 200
(1000 V range)
1200, 200
(1000 V range) *
DC voltage + AC voltage
1200, 200
(1000 V range)
1200, 200
(1000 V range)
Continuity check 100 5900
Diode test 80 3520
Resistance
1200, 120
(60 M
Ω
/600 M
Ω
range)
58800, 5880
(60/600 M
Ω
range)
Conductance 1200 1200
DC
µA 1200 1200
mA 1200 1200
A
1200, 200
(10 A range)
1200, 200
(10 A range)
AC
µA 1200 1200
mA 1200 1200
A
1200, 200
(10 A range)
1200, 200
(10 A range)
AC (clamp input) 5% of the range 5% of the range
* No function is available for the mV range.