92
Accuracy Table
Minimum sensitivity voltage (During DC+ACV measurement, it indicates
an RMS of the AC component.)
Range
[Hz]
AC voltage range (sine wave)
60.000 mV 600.00 mV 6.0000 V 60.000 V 600.00 V 1000.0 V
99.999
*2,*3
6.000 mV 60.00 mV 0.6000 V 6.000 V 60.00 V 60.0 V
999.99
6.000 mV 60.00mV 0.6000 V 6.000 V 60.00 V 60.0 V
9.9999 k
6.000 mV 60.00 mV 0.6000 V 6.000 V 60.00 V 60.0 V
99.999 k
6.000 mV 60.00 mV 0.6000 V
6.000 V
Not
specied
Not
specied
500.00 k
20.000 mV 100.00 mV 1.0000 V
Not
specied
Not
specied
Not
specied
The maximum input is within each range. (The 1000.0 V range depends on
the ACV accuracy table)
*2: The minimum sensitivity voltage which is less than 5 Hz is tripled.
*3: With the lter ON, the minimum sensitivity voltage is multiplied by 0.7.
Minimum sensitivity current
Range
[Hz]
AC range (sine wave)
600.00 µA 6000.0 µA 60.000 mA 600.00 mA 6.0000 A 10.000 A
99.999
*4
60.00 µA 600.0 µA 6.000 mA 60.00 mA 0.6000 A 4.000 A
999.99
60.00 µA 600.0 µA 6.000 mA 60.00 mA 0.6000 A 4.000 A
9.9999 k
60.00 µA 600.0 µA 6.000 mA 60.00 mA 0.6000 A 4.000 A
99.999 k
60.00 µA 600.0 µA 6.000 mA
*6
60.00 mA
*6
Not
specied
Not
specied
500.00 k
100.00 µA
*5
1000.0 µA
*5
Not
specied
Not
specied
Not
specied
Not
specied
The maximum input is within each range.
*4: The minimum sensitivity current which is less than 5 Hz is tripled.
*5: Specied in 200 kHz or less.
*6: Specied in 30 kHz or less.
HIOKI DT4281A981-08