3.2 Making Settings on the Logger
31
3
3
Chapter 3 Settings
The power save function turns off the display 30 seconds after the last button is
pressed, except when the recording interval setting is displayed. The display reap-
pears upon the next button press.
See: "Appendix 3 Battery Life Approximation" (p.A2)
Recording Mode Setting
1
Press the SET button to display the
recording mode setting.
(The [STAT] indica-
tor appears, and the setting blinks.)
2
Press the (+) and (-) buttons to select
[ON] or [OFF].
3
Press the SET button to accept the setting.
(The power save setting is displayed.)
Setting Options
Description
OFF (Default)
The instantaneous value is recorded at each recording interval (instantaneous
recording).
ON
When on, measurements are taken once per second, and instantaneous, maxi-
mum, minimum, and average values are recorded at each recording interval. (sta-
tistical recording).
(Up to 15,000 data values can be recorded.)
Statistical recording cannot be selected when the recording interval is
set to one second.
Power Save Setting
1
Press the SET button to display the power
save setting
([APS] appears, and the setting
blinks).
2
Press the (+) and (-) buttons to select
[ON] or [OFF].
3
Press the SET button to accept the set-
ting.
(The measurement display appears.)
Setting Options
Description
ON (Default) Power save is enabled.
OFF Power save is disabled (the display remains visible).
The Auto Power Save feature consumes a small amount of current