EasyManua.ls Logo

Hirschmann PAT IFLEX5 - Zero-Setting Transducer Inputs; Zero-Setting Slew Potentiometer

Hirschmann PAT IFLEX5
79 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Screen For Sensor Calibration
© Hirschmann Rev. I 05/22/17 190154_I.DOC
55
17.2 ZERO-SETTING THE TRANSDUCER INPUTS
NOTE: The only thing adjustable for the pressure transducers is the zero point, which is the voltage
the transducer outputs when there is no (zero) pressure sensed.
CAUTION: Ensure there is no pressure in the hydraulic line when disconnecting the hoses from
pressure transducers.
The display shows which transducer (piston-
side, rod-side or force) is being zeroed and a
horizontal dial marks the present pressure (or
force) difference in %. NOTE: there are 2
screens piston and rod transducers. By pressing
the + key, the input pressure (or force) is
adjusted upwards, and by pressing the minus (-)
key, the input value is adjusted downwards.
When the plus (+) and minus (-) keys are
pressed simultaneously, the zero setting occurs
automatically. Manual adjustments may be
preformed using + or -.
17.3 ZERO-SETTING THE SLEWING INPUTS
Now, having successfully entered a valid
password, the piston-side zero-point setting
function is activated. Press enter through the
piston and rod zero adjustment screens to the
slew zero-point adjustment function is activated.
An indicator line shows the current position of
the slew pot signal. Note: the indicator line must
be shown in the display scale from -10° to +10°
for this zeroing procedure.
By pressing the two center keys simultaneously,
the zero setting occurs automatically. Note: The
indicator line will move to zero on the bar graph.
When the operator is finished, pressing the EXC or INFO key returns the console display to normal.

Table of Contents

Related product manuals