Service Screen For Sensor Calibration 
© Hirschmann Rev. I 05/22/17  190154_I.DOC 
17.2  ZERO-SETTING THE TRANSDUCER INPUTS 
 
NOTE: The only thing adjustable for the pressure transducers is the zero point, which is the voltage 
the transducer outputs when there is no (zero) pressure sensed. 
 
CAUTION: Ensure there is no pressure in the hydraulic line when disconnecting the hoses from 
pressure transducers. 
 
The display shows which transducer (piston-
side, rod-side or force) is being zeroed and a 
horizontal dial marks the present pressure (or 
force) difference in %. NOTE: there are 2 
screens piston and rod transducers. By pressing 
the + key, the input pressure (or force) is 
adjusted upwards, and by pressing the minus (-) 
key, the input value is adjusted downwards. 
When the plus (+) and minus (-) keys are 
pressed simultaneously, the zero setting occurs 
automatically.  Manual adjustments may be 
preformed using + or -. 
 
 
17.3  ZERO-SETTING THE SLEWING INPUTS 
Now, having successfully entered a valid 
password, the piston-side zero-point setting 
function is activated. Press enter through the 
piston and rod zero adjustment screens to the 
slew zero-point adjustment function is activated. 
An indicator line shows the current position of 
the slew pot signal. Note: the indicator line must 
be shown in the display scale from -10° to +10° 
for this zeroing procedure. 
By pressing the two center keys simultaneously, 
the zero setting occurs automatically. Note: The 
indicator line will move to zero on the bar graph. 
 
 
When the operator is finished, pressing the EXC or INFO key returns the console display to normal.