Step Touchscreen operation Screen displays and notes
5 b-1) To confirm registered
data, press the [Folder]
button you want to verify,
and press the [MENU]
button after the folder
turns blue.
b-2) Press the [Reference]
button in the [MENU]
field.
Press the [Select] button
if the target program
appears.
If the progr
am that appeared is not the program you wanted,
press the [X] button to return to the Program list screen and
check a different folder.
6 Do not change the run
parameters after calling the
program.
Operate the centrifuge according to the instructions in
Section 3-2-5 "Operation Procedures."
Reentering (changing) the run parameters (for example,
SPEED, TIME, etc.) after calling a program will cancel a
called program.
To combine programmed operation and RTC operation (refer to Section 3-3-4 RTC (Real
Time Control) operation"), set RTC after calling the program. When a program includes
step-mode operation, the run time of all steps must be calculated to compute RTC start time
since the program cannot be called once RTC is set.
The MENU box appears
in the Program list
screen.
[Select] button
3-3-1-2 Programmed Operation Procedures
[Reference] button
MENU box
The Check registered
parameter screen
appears.
NOTE
3-40