3 - 15
[Reference]
Auto parameter setting (fixed)
<Sensitivity>
Wavelength scan (acquisition of signal intensity)
EX and EM Slits : Both 10 nm
EX WL : 350 nm
EM WL : Raman peak wavelength
Response : 2 s
Scan speed : 60 nm/min
PMT Voltage : 700 V
Time scan (acquisition of noise level)
EX and EM Slits : Both 10 nm
EX WL : 350 nm
EM WL : Raman peak wavelength
Response : 4 s
Scan time : 600 s
PMT Voltage : 700 V
<Sensitivity [BG]>
Condition setting Wavelength scan
EX and EM Slits : Both 10 nm
EX WL : 350 nm
EM WL : Peak WL of Raman
Response : 4 s
Speed : 60 nm/min
PMT Voltage : 700 V