3 - 27
3.6.2 Error Message List
Table 3-2 indicates the error information that may appear on the screen of the PC. If any of the errors
shown here appear, take a remedy as indicated in the table.
Table 3-2 Error Message List
Contact our local service office.
Contact our local service office.
Lamp extinguished during
initialization.
Momentary power failure
detected at instrument.
Check the connection of USB cable, and
start up PC and spectrophotometer again.
If error recurs, contact our local service
office.
Contact our local service office.
Auto Pre-scan error!
Peak not Found!
Peak was not detected in auto
pre-scan.
Peak may be detected by widening the slit
or increasing photomultiplier voltage.
Pre-scan error!
Peak not Found!
Peak was not detected in normal
pre-scan.
Peak may be detected by widening the slit
or increasing photomultiplier voltage.
Wavelength
Initialization error
(EX)!
Initialize position of EX side
wavelength feed screw cannot be
detected.
Contact our local service office.
Wavelength
Initialization error
(EM)!
Initialize position of EM side
wavelength feed screw cannot be
detected.
Contact our local service office.
Shutter Initialization
error
Initialize position of shutter
cannot be detected.
Contact our local service office.
EX Slit initialization
error!
Initialize position of EX side slit
cannot be detected.
Contact our local service office.
EM Slit initialization
error!
Initialize position of EM side slit
cannot be detected.
Contact our local service office.
Sync signal from chopper cannot
be detected.
Contact our local service office.
Abnormality detected in
spectrophotometer main unit
program.
Contact our local service office.
The Xe lamp was
turned off suddenly.
Turn off the main
switch of the
spectrophotometer
unit.
Replace with a new Xe lamp.