8-459. 
Executive and HP-IB Block Signature Analysis Troubleshooting 
8-460.  Signature Analysis 
can 
be used to troubleshoot the Executive and HP-IB Block components. Follow the 
instructions given below. 
Setup: 
a. 
Begin with the 
5334B in STANDBY. 
b. 
Connect a jumper between the Executive MCU 
(U19) ST or SP pin (TP17 or TP18) and Ground 
(TF'8). 
c. 
Connect a jumper  between the HP-IB MCU 
(U17) ST or SP pin (TP14 or TP 15) and Ground (TP8). 
d. 
Apply power to the 5334B and observe that the Counter displays "SA 
dIAG. 
e.  Remove jumpers from ST/SP and Ground for both MCUs. 
NOTE 
The 
53348 
must 
remain 
ON 
and displaying 
"SA 
diAG" 
f. 
To check the EXECUTIVE BLOCK using Signature Analysis, connect Pod Leads from the signature 
analyzer to the 
A1 Main Board as follows: 
Signature 
Analyzer 
U19 
Test Points 
START LEAD 
ST 
(TP17) 
STOP LEAD 
SP 
(TP18) 
CLOCK LEAD 
SA CLK 
(TF'13) 
GROUND LEAD  Ground (TP8) 
1. 
Apply power to the signature analyzer and observe that its GATE light is flashing. 
2. 
Set the signature analyzer push button switches as follows: 
START BUTTON  Rising Edge 
STOP BUTTON  Falling Edge 
CLOCK BUTTON 
Falling 
Edge 
3. 
Take signatures. Signatures should match those shown in 
Figure 
8-13. 
HP 
5334B 
- 
Service Manual 
8-76