8-470.
SIGNATURE ANALYSIS SETUP
#3
8-471. To perform signature analysis for Setup
#3,
proceed as follows:
a.
Switch off power to the Counter.
b.
Attach the test pod leads of the signature analyzer
as
follows:
STISPIISTART: 5701, pin
6
QUAW/STOP: U709, pin
22
CLOCK: 5701, pin
7
(CLK)
GND: 5701, pin
9
(GND)
c.
Set the signature analyzer
as
follows:
FUNCTION. QUAL
CLOCK: falling edge
START: rising edge
STOP: rising edge
QUAL: low level
d.
Power up the Counter and verify the signatures shown in
Figure
8-17.
HP
5334B
-
Service Manual
8-87