5.
Set the HP
5334B
as follows:
CHAN
A
Negative SLOPE
................................................................................
ON (falling edge)
VERIFY:
The Counter displays
30
ns
+
10
ns (Fall Time).
6.
Record the fall measurement on the Performance Test Record Card, lime
39.
7.
Set the function generator as follows:
Frequency
...............................................................................................................
100
Hz
Amplitude
..............................................................................................................
1v
P-P
Function
.................................................................................................................
Negative Ramp
VERIFY:
The Counter displays
8
ms
2
0.6
ms (Fall Time).
8.
Record the fall measurement on the Performance Test Record Card, line
40.
9.
Set the function generator as follows:
Function
................................................................................................................. Positive Ramp
10.
Set the HP
5334B
as follows:
CHAN
A
Negative SLOPE
.................................................................................
OFF (rising edge)
VERIFY:
The Counter displays
8
ms
+
0.6
ms (Rise Time).
11.
Record the rise measurement on the Performance Test Record Card, line
41.
Failure:
If the instrument under test does not meet the test specification, consider performing the adjustments in
Section
V
of this manual as a first step
in
correcting the problem.
HP
5334B
-
Service Manual
4-33